Nanjing Perlong Medical Equipment Co., Ltd
|
|
| |
| Products |
|
 |
XFA6000 Auto Hematology Analyzer |
|
|
|

|

|
Click to enlarge image
|
Model Number |
XFA6000 |
| XFA6000 Auto Hematology Analyzer |
|
| Specifications |
| Test Items |
WBC, LY#, MO#, GR#, LY%, MO%, GR%, HGB, RBC, HCT, MCV, MCH, MCHC, RDW?V, RDW?D, PLT, PCT, MPV, PDW, WBC Histogram, RBC Histogram, PLT Histogram |
| Background Count |
WBC: ≤0.3?09/L RBC: ≤0.03?012/L HGB: ≤1g/L PLT: ≤10?09/L |
| Linearity |
Test Item |
Test Range |
Allow Difference of Linearity |
| WBC |
0∼6.0?09/L |
±0.3?09/L |
| 6.0∼99.9?09/L |
±5% |
| RBC |
0∼0.99?012/L |
±0.05?012/ L |
| 1.0∼9.99?012/L |
±5% |
| HGB |
0∼99 g/L |
±2g/L |
| 100∼300 g/L |
±2% |
| PLT |
0∼99?09/L |
±10?09/L |
| 100∼999?09/L |
±10% |
| Repeatability |
Test Item |
CV(%) |
Teat Range |
| WBC |
≤2.0 |
7.0ï½15.0?09/L |
| RBC |
≤0.5 |
3.5ï½6.0?012/L |
| HGB |
≤1.5 |
110ï½180g/L |
| PLT |
≤4.0 |
150ï½500?09/L |
| MCV |
≤0.5 |
80ï½110fL |
| Sample Volume |
Twig blood: 20μl |
| Dilution Ratio |
WBC/HGB: 1:500 RBC/PLT: 1:50000 |
| Interfaces |
RS232 joint, USB, PS2, Parallel port |
| Test Speed |
60 samples/hours |
| Continuous Work Time |
>8h |
|
|
|
|
|
|
| |
|
|
|